Dräger Tube n-butanol 10/a - PN: 8103861
Rated 3.5/5 based on 11 customer reviews

Dräger Tube n-butanol 10/a - PN: 8103861

$104.00
Image

Dräger Tube n-butanol 10/a (10)
PN: 8103861

Application Range:

  • Standard Measuring Range: 10 to 250 ppm
  • Number of Strokes n: 20
  • Time for Measurement: Approx. 6 mins
  • Standard Deviation: ± 10 to 25 %
  • Color Change: Yellow → Mint Green

Ambient Operating Conditions:

  • Temperature: 15 to 30 °C
  • Absolute Humidity: 3 to 15 mg H2O / L

Reaction Principle:

  • n-butanol + organometallic compound → green reaction product

Cross Sensitivity:

  • The tube does not differentiate between different alcohols. 2-butanol is indicated with the same sensitivity. During the measurement of isobutanol with n = 2/10 strokes, the concentration read must be multiplied by a factor of 0.4. During the measurement of tert-butanol with n = 2/10 strokes, the concentration read must be multiplied by a factor of 3.0.
  • Methanol is indicated with 2 times (n=10) to 3 times (n=2) its sensitivity, ethanol and isopropanol are indicated with 1 time (n=10) to 2 times (n=2) their sensitivity. Higher molecular alcohols are indicated with significantly decreasing sensitivity. Ethers are indicated with a different sensitivity. ≤ 25 ppm formaldehyde, ≤ 50 ppm acetaldehyde, and ≤ 50 ppm toluene are not indicated.
  • Aliphatic petroleum hydrocarbons, ketones, esters, halogenated hydrocarbons and benzene are not indicated.

Dräger Tube n-butanol 10/a (10)
PN: 8103861

Application Range:

  • Standard Measuring Range: 10 to 250 ppm
  • Number of Strokes n: 20
  • Time for Measurement: Approx. 6 mins
  • Standard Deviation: ± 10 to 25 %
  • Color Change: Yellow → Mint Green

Ambient Operating Conditions:

  • Temperature: 15 to 30 °C
  • Absolute Humidity: 3 to 15 mg H2O / L

Reaction Principle:

  • n-butanol + organometallic compound → green reaction product

Cross Sensitivity:

  • The tube does not differentiate between different alcohols. 2-butanol is indicated with the same sensitivity. During the measurement of isobutanol with n = 2/10 strokes, the concentration read must be multiplied by a factor of 0.4. During the measurement of tert-butanol with n = 2/10 strokes, the concentration read must be multiplied by a factor of 3.0.
  • Methanol is indicated with 2 times (n=10) to 3 times (n=2) its sensitivity, ethanol and isopropanol are indicated with 1 time (n=10) to 2 times (n=2) their sensitivity. Higher molecular alcohols are indicated with significantly decreasing sensitivity. Ethers are indicated with a different sensitivity. ≤ 25 ppm formaldehyde, ≤ 50 ppm acetaldehyde, and ≤ 50 ppm toluene are not indicated.
  • Aliphatic petroleum hydrocarbons, ketones, esters, halogenated hydrocarbons and benzene are not indicated.