Dräger Tube Mercury Vapor 0.1/b - PN: CH23101
Dräger Tube Mercury Vapor 0.1/b (10)
PN: CH23101
Application Range
Standard Measuring Range: 0.05 to 2 mg/m3
Number of Strokes n: 40 to 1
Time for Measurement: max. 10 min
Standard Deviation: ± 30 %
Color Change: pale yellow grey ä pale orange
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: < 20 mg H2O / L
Reaction Principle
Hg + Cul ä Cu-Hg-complex
Cross Sensitivity
Free halogens cause substantial minus errors. It is impossible to measure mercury vapor in the presence of halogens. Arsine, phosphine, hydrogen sulfide, ammonia, nitrogen dioxide, sulfur dioxide and hydrazine in the TLV range do not interfere.