Application Range
Standard Measuring Range: 30 to 400 ppm
Number of Strokes n: 6
Time for Measurement: approx. 2 min
Standard Deviation: ± 10 to 15 %
Color Change: white ä brown

Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: 5 to 12 mg H2O / L

Reaction Principle
C6H5 - C2H5 + I2O5 ä I 2

Cross Sensitivity
Many petroleum hydrocarbons and aromatics are indicated, but with different sensitivities. It is impossible to differentiate between them. Extension of the Measuring Range Using n = 4, multiply the reading by 1.5; the measuring range will be 45 to 600 ppm.