Dräger Tube Chloroform 2/a - PN: 6728861
Dräger Tube Chloroform 2/a (5)
PN: 6728861
Application Range
Standard Measuring Range: 2 to 10 ppm
Number of Strokes n: 10
Time for Measurement: approx. 9 min
Standard Deviation: ± 20 to 30 % at 20 °C and 9 mg H2O / L
Color Change: white ä yellow
Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: 9 mg H2O / L
Reaction Principle
a) CHCl3 + CrVI ä Cl2
b) Cl2 + o-tolidine ä yellow reaction product
Cross Sensitivity
Other chlorinated hydrocarbons are indicated, but with different sensitivities.
Dräger Tube Chloroform 2/a (5)
PN: 6728861
Application Range
Standard Measuring Range: 2 to 10 ppm
Number of Strokes n: 10
Time for Measurement: approx. 9 min
Standard Deviation: ± 20 to 30 % at 20 °C and 9 mg H2O / L
Color Change: white ä yellow
Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: 9 mg H2O / L
Reaction Principle
a) CHCl3 + CrVI ä Cl2
b) Cl2 + o-tolidine ä yellow reaction product
Cross Sensitivity
Other chlorinated hydrocarbons are indicated, but with different sensitivities.