Dräger Tube Carbon Tetrachloride 1/a - PN: 8101021
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Carbon Tetrachloride 1/a - PN: 8101021

$108.00
Image

Dräger Tube Carbon Tetrachloride 1/a (10)
PN: 8101021

Application Range
Standard Measuring Range: 1 to 15 ppm / 10 to 50 pmm
Number of Strokes n: 10 / 5
Time for Measurement: approx. 6 min / 3 min
Standard Deviation: ± 15 to 20 %
Color Change: white ä yellow

Ambient Operating Conditions
Temperature: 15 to 30 °C
Absolute Humidity: 3 to 15 mg H2O / L

Reaction Principle
a) CCI4 + H2S2O7 ä COCI2
b) COCI2 + Aromatic nitrogen compounds ä yellow reaction product

Cross Sensitivity
Chloropicrin and phosgene are indicated with the same sensitivity. It is impossible to measure carbon tetrachloride in the presence of chloropicrin and phosgene.
No interference from:
1 ppm chlorine
5 ppm hydrochloric acid
20 ppm methyl bromide
1,000 ppm acetone

Dräger Tube Carbon Tetrachloride 1/a (10)
PN: 8101021

Application Range
Standard Measuring Range: 1 to 15 ppm / 10 to 50 pmm
Number of Strokes n: 10 / 5
Time for Measurement: approx. 6 min / 3 min
Standard Deviation: ± 15 to 20 %
Color Change: white ä yellow

Ambient Operating Conditions
Temperature: 15 to 30 °C
Absolute Humidity: 3 to 15 mg H2O / L

Reaction Principle
a) CCI4 + H2S2O7 ä COCI2
b) COCI2 + Aromatic nitrogen compounds ä yellow reaction product

Cross Sensitivity
Chloropicrin and phosgene are indicated with the same sensitivity. It is impossible to measure carbon tetrachloride in the presence of chloropicrin and phosgene.
No interference from:
1 ppm chlorine
5 ppm hydrochloric acid
20 ppm methyl bromide
1,000 ppm acetone