Dräger Tube Carbon Disulfide 30/a - PN: CH23201
Dräger Tube Carbon Disulfide 30/a (10)
PN: CH23201
Application Range
Standard Measuring Range: 0.1 to 10 mg/L
Number of Strokes n: 6
Time for Measurement: approx. 1 min
Standard Deviation: ± 15 to 20 %
Color Change: pale blue ä brown
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: < 30 mg H2O / L
Reaction Principle
2 CS2 + 4 NHR2 + Cu2+ ä Cu(SCSNR2)2 + 2 NH2R2 +
Cross Sensitivity
Hydrogen sulfide is indicated, producing a pale green discoloration. It is impossible to measure carbon disulfide in the presence of hydrogen sulfide.