Dräger Tube Acetone 100/b - PN: CH22901
Dräger Tube Acetone 100/b (10)
PN: CH22901
Application Range
Standard Measuring Range: 100 to 12,000 ppm
Number of Strokes n: 10
Time for Measurement: approx. 4 min
Standard Deviation: ± 15 to 20 %
Color Change: pale yellow ä yellow
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: 5 to 20 mg H2O / L
Reaction Principle
Acetone + 2.4-Dinitrophenylhydrazine ä yellow hydrazone
Cross Sensitivity
Other ketones are indicated but with different sensitivities. Aldehydes are indicated, but not esters. Ammonia causes the indicating layer to turn yellowish brown.
Dräger Tube Acetone 100/b (10)
PN: CH22901
Application Range
Standard Measuring Range: 100 to 12,000 ppm
Number of Strokes n: 10
Time for Measurement: approx. 4 min
Standard Deviation: ± 15 to 20 %
Color Change: pale yellow ä yellow
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: 5 to 20 mg H2O / L
Reaction Principle
Acetone + 2.4-Dinitrophenylhydrazine ä yellow hydrazone
Cross Sensitivity
Other ketones are indicated but with different sensitivities. Aldehydes are indicated, but not esters. Ammonia causes the indicating layer to turn yellowish brown.