Dräger MT Formaldehyde 0.15-3 ppm - PN 8610100
Rated 3.5/5 based on 11 customer reviews

Dräger MT Formaldehyde 0.15-3 ppm - PN 8610100

$105.00
Image

Dräger MT Formaldehyde 0.15-3 ppm
PN : 8610100

 
Measuring Range
0.15 to 3 ppm
Typical Measuring Time 
300 to 600 s
Accuracy  ± 25% ( under calibration conditions )
Ambient Conditions

Temperature 
0 to 40 °C
Humidity 
4 to 20 mg/L
Humidity correction  <8 mg/L to 40% from measurement value per mg/L
8 to 12 mg/L to none 
>12 mg/L to 10%
from measurement value per mg/L
Cross Sensitivity

No influence at 0.6 ppm formaldehyde by:
0.25 ppm acrolein
10 ppm vinyl acetate
50 ppm acetaldehyde
100 ppm styrene
formaldehyde  measurement is not possible in the presence of SO2

Hose and probe measurement 
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose  and probes.



 Hose 
Length ( meter )
Flashing time (min)
FKM hose 
5 - 20 
3
FKM hose 
45
5
PTFE-line PVC hose
3 - 15 3
Bar probe 90 (83 16 532)

1
Telescopic probe ES 150 ( 83 16 533)

2
Probe set GP600 )83 28 667)
3



Dräger MT Formaldehyde 0.15-3 ppm
PN : 8610100

 
Measuring Range
0.15 to 3 ppm
Typical Measuring Time 
300 to 600 s
Accuracy  ± 25% ( under calibration conditions )
Ambient Conditions

Temperature 
0 to 40 °C
Humidity 
4 to 20 mg/L
Humidity correction  <8 mg/L to 40% from measurement value per mg/L
8 to 12 mg/L to none 
>12 mg/L to 10%
from measurement value per mg/L
Cross Sensitivity

No influence at 0.6 ppm formaldehyde by:
0.25 ppm acrolein
10 ppm vinyl acetate
50 ppm acetaldehyde
100 ppm styrene
formaldehyde  measurement is not possible in the presence of SO2

Hose and probe measurement 
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose  and probes.



 Hose 
Length ( meter )
Flashing time (min)
FKM hose 
5 - 20 
3
FKM hose 
45
5
PTFE-line PVC hose
3 - 15 3
Bar probe 90 (83 16 532)

1
Telescopic probe ES 150 ( 83 16 533)

2
Probe set GP600 )83 28 667)
3