Dräger Tube Mercury Vapor 0.1/b - PN: CH23101
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Mercury Vapor 0.1/b - PN: CH23101

$107.00
Image

Dräger Tube Mercury Vapor 0.1/b (10)
PN: CH23101

Application Range
Standard Measuring Range: 0.05 to 2 mg/m3
Number of Strokes n: 40 to 1
Time for Measurement: max. 10 min
Standard Deviation: ± 30 %
Color Change: pale yellow grey ä pale orange

Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: < 20 mg H2O / L

Reaction Principle
Hg + Cul ä Cu-Hg-complex

Cross Sensitivity
Free halogens cause substantial minus errors. It is impossible to measure mercury vapor in the presence of halogens. Arsine, phosphine, hydrogen sulfide, ammonia, nitrogen dioxide, sulfur dioxide and hydrazine in the TLV range do not interfere.

Dräger Tube Mercury Vapor 0.1/b (10)
PN: CH23101

Application Range
Standard Measuring Range: 0.05 to 2 mg/m3
Number of Strokes n: 40 to 1
Time for Measurement: max. 10 min
Standard Deviation: ± 30 %
Color Change: pale yellow grey ä pale orange

Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: < 20 mg H2O / L

Reaction Principle
Hg + Cul ä Cu-Hg-complex

Cross Sensitivity
Free halogens cause substantial minus errors. It is impossible to measure mercury vapor in the presence of halogens. Arsine, phosphine, hydrogen sulfide, ammonia, nitrogen dioxide, sulfur dioxide and hydrazine in the TLV range do not interfere.