Dräger Tube Trichlorethylene 2/a - PN: 6728541
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Trichlorethylene 2/a - PN: 6728541

$97.50
Image

Dräger Tube Trichlorethylene 2/a (10)
PN: 6728541

Application Range
Standard Measuring Range: 20 to 250 ppm / 2 to 50 ppm
Number of Strokes n: 3 / 5
Time for Measurement: approx. 1.5 min / 2.5 min
Standard Deviation: ± 10 to 15 %
Color Change: pale grey ä orange

Ambient Operating Conditions
Temperature: 10 to 40 °C
Absolute Humidity: 5 to 15 mg H2O / L

Reaction Principle
Cl2 + o-tolidine ä orange reaction product

Cross Sensitivity
Other chlorinated hydrocarbons are indicated, but with different sensitivities.
Free halogens and hydrogen halides in the TLV range are also indicated. It is impossible to measure trichloroethylene in the presence of these substances. Petroleum hydrocarbons cause low readings.