Dräger-Tube Sulfuric Acid 1/a - PN: 6728781
Rated 3.5/5 based on 11 customer reviews

Dräger-Tube Sulfuric Acid 1/a - PN: 6728781

$139.00
Image

Dräger-Tube Sulfuric Acid 1/a
PN: 6728781



Application Range
Standard Measuring Range 1 to 5 mg /m3 
Number of Strokes n 100
Time for Measurement approx. 100 min 
Standard Deviation ± 30%
Color Change brown to pink violet 
Ambient Operating Conditions
Temperature
5 to 40 °C
Absolute Humidity <15 mg H2O/L 
 Cross Sensitivity
Sulfur trioxide in the gaseous form is not indicated, but in the presence of atmospheric humidity  this forms sulfuric acid aerosols, which are indicated. Soluble sulfates and other aerosol forming acids are also indicated but with different sensitivities . It is impossible to measure sulfuric acid in the presence of these substances.
Additional Information 
After performing the required 100 pump strokes the reagent ampoule must be broken and the liquid completely onto the brown reagent layer. Wait one minute before drawing the liquid carefully through the brown layer with approx. 1/4 pump stroke into the indication chamber. The measurement must be evaluated immediately

Dräger-Tube Sulfuric Acid 1/a
PN: 6728781



Application Range
Standard Measuring Range 1 to 5 mg /m3 
Number of Strokes n 100
Time for Measurement approx. 100 min 
Standard Deviation ± 30%
Color Change brown to pink violet 
Ambient Operating Conditions
Temperature
5 to 40 °C
Absolute Humidity <15 mg H2O/L 
 Cross Sensitivity
Sulfur trioxide in the gaseous form is not indicated, but in the presence of atmospheric humidity  this forms sulfuric acid aerosols, which are indicated. Soluble sulfates and other aerosol forming acids are also indicated but with different sensitivities . It is impossible to measure sulfuric acid in the presence of these substances.
Additional Information 
After performing the required 100 pump strokes the reagent ampoule must be broken and the liquid completely onto the brown reagent layer. Wait one minute before drawing the liquid carefully through the brown layer with approx. 1/4 pump stroke into the indication chamber. The measurement must be evaluated immediately