Dräger Tube Styrene 10/a - PN: 6723301
Dräger Tube Styrene 10/a (10)
PN: 6723301
Application Range
Standard Measuring Range: 10 to 200 ppm
Number of Strokes n: max. 15
Time for Measurement: max. 3 min
Standard Deviation: ± 15 to 20 %
Color Change: white ä pale yellow
Ambient Operating Conditions
Temperature: 15 to 40 °C
Absolute Humidity: < 15 mg H2O / L
Reaction Principle
C6H5-CH=CH2 + H2SO4 ä pale yellow reaction product
Cross Sensitivity
Other organic compounds which tend toward polymerization (e. g. buta diene) are indicated, but with different sensitivities. It is impossible to measure monostyrene in the presence of these compounds.
Dräger Tube Styrene 10/a (10)
PN: 6723301
Application Range
Standard Measuring Range: 10 to 200 ppm
Number of Strokes n: max. 15
Time for Measurement: max. 3 min
Standard Deviation: ± 15 to 20 %
Color Change: white ä pale yellow
Ambient Operating Conditions
Temperature: 15 to 40 °C
Absolute Humidity: < 15 mg H2O / L
Reaction Principle
C6H5-CH=CH2 + H2SO4 ä pale yellow reaction product
Cross Sensitivity
Other organic compounds which tend toward polymerization (e. g. buta diene) are indicated, but with different sensitivities. It is impossible to measure monostyrene in the presence of these compounds.