Dräger Tube Nitrous Fumes 2/a - PN: CH31001
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Nitrous Fumes 2/a - PN: CH31001

$97.50
Image

Dräger Tube Nitrous Fumes 2/a (10)
PN: CH31001

Application Range
Standard Measuring Range: 5 to 100 / 2 to 50 ppm
Number of Strokes n: 5 / 10
Time for Measurement: approx. 1 min / approx. 2 min
Standard Deviation: ± 10 to 15 %
Color Change: yellow ä blue grey

Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: max. 30 mg H2O / L

Reaction Principle
a) NO + CrVI ä NO2
b) NO2 + o-diphenylbenzidine ä blue grey reaction product

Cross Sensitivity
Chlorine and ozone are indicated with different sensitivities.

Dräger Tube Nitrous Fumes 2/a (10)
PN: CH31001

Application Range
Standard Measuring Range: 5 to 100 / 2 to 50 ppm
Number of Strokes n: 5 / 10
Time for Measurement: approx. 1 min / approx. 2 min
Standard Deviation: ± 10 to 15 %
Color Change: yellow ä blue grey

Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: max. 30 mg H2O / L

Reaction Principle
a) NO + CrVI ä NO2
b) NO2 + o-diphenylbenzidine ä blue grey reaction product

Cross Sensitivity
Chlorine and ozone are indicated with different sensitivities.