Dräger-Tube Hydrochloric Acid/Nitric Acid 1/a - PN: 8101681
Dräger-Tube Hydrochloric Acid/Nitric Acid 1/a - PN: 8101681
$133.00
Dräger-Tube Hydrochloric Acid/Nitric Acid 1/a (10)
PN: 8101681
Application Range | |
Standard Measuring Range | 1 to 10 ppm / 1 to 15 ppm |
Number of Strokes n | 10 / 20 |
Time for Measurement | approx. 1.5 min / approx. 3 min |
Standard Deviation | ± 30% |
Color Change | blue to yellow |
Ambient Operating Conditions | |
Temperature |
5 to 40 °C for HCl |
For HNO3 measurements, the tube scales are applicable only to 20 °C. Where the temperature is different, the measuring result must be multiplied as follows: | |
Temperature Factor | |
40 0.3 30 0.4 10 2 |
|
Absolute Humidity | max 15 mg H2O/L |
Cross Sensitivity |
50 ppm nitrogen dioxide result in almost the same reading as 2 ppm nitric acid. 10 ppm hydrogen sulfide or 5 ppm nitrogen dioxide do not affect the reading. Chlorine concentrations in excess of 1 ppm change the entire indicating layer to yellow-green |
Dräger-Tube Hydrochloric Acid/Nitric Acid 1/a (10)
PN: 8101681
Application Range | |
Standard Measuring Range | 1 to 10 ppm / 1 to 15 ppm |
Number of Strokes n | 10 / 20 |
Time for Measurement | approx. 1.5 min / approx. 3 min |
Standard Deviation | ± 30% |
Color Change | blue to yellow |
Ambient Operating Conditions | |
Temperature |
5 to 40 °C for HCl |
For HNO3 measurements, the tube scales are applicable only to 20 °C. Where the temperature is different, the measuring result must be multiplied as follows: | |
Temperature Factor | |
40 0.3 30 0.4 10 2 |
|
Absolute Humidity | max 15 mg H2O/L |
Cross Sensitivity |
50 ppm nitrogen dioxide result in almost the same reading as 2 ppm nitric acid. 10 ppm hydrogen sulfide or 5 ppm nitrogen dioxide do not affect the reading. Chlorine concentrations in excess of 1 ppm change the entire indicating layer to yellow-green |