Dräger Sulphur Dioxide 0.05 - 5 ppm - PN 8610110
Dräger Sulphur Dioxide 0.05 - 5 ppm
PN : 8610110
Measuring Range |
0.05 to 5 ppm |
Typical Measuring Time |
20 to 480 s |
Accuracy | ± 25% ( under calibration conditions ) |
Ambient Conditions |
|
Temperature |
0 to 30 °C |
Humidity |
1 to 20 mg/L |
Cross Sensitivity |
No influence at 0.5 ppm sulfur dioxide by: 150 ppm hydrogen sulfide 5 ppm hydrogen chlorine |
Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
Hose |
Length ( meter ) |
Flashing time (min) |
FKM hose |
5 - 20 |
10 |
FKM hose |
45 |
20 |
PTFE-line PVC hose |
3 - 15 | 3 |
Bar probe 90 (83 16 532) |
3 | |
Telescopic probe ES 150 ( 83 16 533) |
15 | |
Probe set GP600 )83 28 667) | 10 | |
Dräger Sulphur Dioxide 0.05 - 5 ppm
PN : 8610110
Measuring Range |
0.05 to 5 ppm |
Typical Measuring Time |
20 to 480 s |
Accuracy | ± 25% ( under calibration conditions ) |
Ambient Conditions |
|
Temperature |
0 to 30 °C |
Humidity |
1 to 20 mg/L |
Cross Sensitivity |
No influence at 0.5 ppm sulfur dioxide by: 150 ppm hydrogen sulfide 5 ppm hydrogen chlorine |
Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
Hose |
Length ( meter ) |
Flashing time (min) |
FKM hose |
5 - 20 |
10 |
FKM hose |
45 |
20 |
PTFE-line PVC hose |
3 - 15 | 3 |
Bar probe 90 (83 16 532) |
3 | |
Telescopic probe ES 150 ( 83 16 533) |
15 | |
Probe set GP600 )83 28 667) | 10 | |