Dräger DT Hydrogen 0.5%/a (10) - PN CH30901
Rated 3.5/5 based on 11 customer reviews

Dräger DT Hydrogen 0.5%/a (10) - PN CH30901

$98.50
Image

Dräger DT Hydrogen 0.5%/a (10)
PN : CH30901



Application Range
Standard Measuring Range 0.5 to 3.0 vol.%
Number of Strokes n 5
Time for Measurement approx. 1 min 
Standard Deviation ± 10 to 15%
Color Change yellow green to pink
Ambient Operating Conditions
Temperature
5 to 40 °C
Absolute Humidity < 30 mg H2O/L 
 Cross Sensitivity
Up to 1.000 ppm CO has no influence on the indication; higher concentration lead to lower measurement results. Acetylene and alcohols react similarly to hydrogen.
Additional Information 
Do not use in potentially explosive areas. Qualify before use with a combustible gas monitor. When the hydrogen concentration is above 3 vol.% the catalysis layer heats up during the measurement with a reddish glow. Determination of hydrogen in air with at least 5 vol.% O2. 

Dräger DT Hydrogen 0.5%/a (10)
PN : CH30901



Application Range
Standard Measuring Range 0.5 to 3.0 vol.%
Number of Strokes n 5
Time for Measurement approx. 1 min 
Standard Deviation ± 10 to 15%
Color Change yellow green to pink
Ambient Operating Conditions
Temperature
5 to 40 °C
Absolute Humidity < 30 mg H2O/L 
 Cross Sensitivity
Up to 1.000 ppm CO has no influence on the indication; higher concentration lead to lower measurement results. Acetylene and alcohols react similarly to hydrogen.
Additional Information 
Do not use in potentially explosive areas. Qualify before use with a combustible gas monitor. When the hydrogen concentration is above 3 vol.% the catalysis layer heats up during the measurement with a reddish glow. Determination of hydrogen in air with at least 5 vol.% O2.