Dräger Carbon Dioxide 200 - 50000 ppm - PN 8610190
Dräger Carbon Dioxide 200 - 50000 ppm
PN : 8610190
Measuring Range |
200 to 50,000 ppm |
Typical Measuring Time |
2 to 120 s |
Accuracy | ± 25% ( under calibration conditions ) |
Ambient Conditions |
|
Temperature |
0 to 30 °C |
Humidity |
1 to 30 mg/L |
Cross Sensitivity |
No influence at 5000 ppm carbon dioxide by: 100 ppm hydrogen sulfide 200 ppm sulfur dioxide |
Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
Hose |
Length ( meter ) |
Flashing time (min) |
FKM hose |
5 - 20 |
3 |
FKM hose |
45 |
5 |
PTFE-line PVC hose |
3 - 15 | 3 |
Bar probe 90 (83 16 532) |
2 |
|
Telescopic probe ES 150 ( 83 16 533) |
2 |
|
Probe set GP600 )83 28 667) | 3 | |
Dräger Carbon Dioxide 200 - 50000 ppm
PN : 8610190
Measuring Range |
200 to 50,000 ppm |
Typical Measuring Time |
2 to 120 s |
Accuracy | ± 25% ( under calibration conditions ) |
Ambient Conditions |
|
Temperature |
0 to 30 °C |
Humidity |
1 to 30 mg/L |
Cross Sensitivity |
No influence at 5000 ppm carbon dioxide by: 100 ppm hydrogen sulfide 200 ppm sulfur dioxide |
Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
Hose |
Length ( meter ) |
Flashing time (min) |
FKM hose |
5 - 20 |
3 |
FKM hose |
45 |
5 |
PTFE-line PVC hose |
3 - 15 | 3 |
Bar probe 90 (83 16 532) |
2 |
|
Telescopic probe ES 150 ( 83 16 533) |
2 |
|
Probe set GP600 )83 28 667) | 3 | |